Description

C 200

C 200-System consists of the following components:
Measuring cell C 200
Decomposition vessel C 5010
Oxygen filling station C 248
Consumables for calibrations and installation

Space-saving and low cost combustion calorimeter for determing gross calorific values of liquid and solid samples.

Validation according to DIN 51900,ISO 1928, ASTM D240, ASTM D4809, ASTM D5865, ASTM D1989, ASTM D5468, ASTM E711.

Suitable for teaching and training, as well (f.e tecnical- schools, universities) and for industriell laboratories with less analyses..

Manual handling for water and oxygen filling.

Four different working methods, isoperibolic, dynamic, manual, time controlled makes possible an ideal adjustment of different operating tasks.

Clear and easily explained display which is simply to use .

With an external power pack, "world tension useable" from 100 - 240 V AC, 50/60 Hz. The operating voltage of the calorimeter is 24 V DC low- voltage.

Friendly in aftersales

Manual ( teaching mode): ignition and end of the measurement will be done by the operator himself; the temperature increase will be indicated at the display each minute.
All calculations have to be done manually.
At the other three modes ignition and calculation of the gross calorific value will be done automatically.
The gross calorific value will be shown on the display. Acid corrections and calculations of the net calorific value have to be done manually.

The modes have different measuring times:
Isoperibolic: approx: 17 min.
Dynamic : approx. 8 min
Manual : approx. 17 min
(depend on the operator)
Time controlled : 14 min

The decomposition vessel can be equiped in to use an combustible crucibles C 14
(accessory C 5010.4 attachment is neessary)

User-friendly software C 6040 Calwin for controlling the calorimeter and administration of measuring data (asseccory). Transfer of the datas to Excel or Word are possible
Up to eight IKA measuring cells can be controlled by a single PC, using a multiserial plug in card PCI 8.2 (accessory)
Scope of delivery
  • C 200
  • C 5010 Decomposition vessel, standard
  • C 248 Oxygen filling station
  • Ignation adapter
  • C 200.1 Measuring cup 2000 ml
  • Water emptying hose

Technical Data

Measuring range max. 40000 J
Measuring mode dynamic 25°C yes
Measuring mode isoperibol 25°C yes
Measuring time dynamic approx. 8 min
Measuring time isoperibol approx. 17 min
Reproducibility dynamic (1g benzoic acid NBS39i) 0.1 %RSD
Reproducibility isoperibol (1g benzoic acid NBS39i) 0.1 %RSD
Working temperature max. 25 °C
Temperature measurement resolution 0.0001 K
Interface printer Centronix
Interface PC RS232
Decomposition vessel C 5010 yes
Works according to DIN 51900 yes
Works according to DIN EN ISO 1716 yes
Works according to DIN EN ISO 18125 yes
Works according to DIN EN 15400 yes
Works according to DIN CEN TS 14918 yes
Works according to DIN CEN/TS 16023 yes
Works according to DIN SPEC 19524 yes
Works according to ASTM D240 yes
Works according to ASTM D4809 yes
Works according to ASTM D5468 yes
Works according to ASTM D5865 yes
Works according to ISO 1928 yes
Works according to GOST Certified yes
Dimensions (W x H x D) 400 x 400 x 400 mm
Weight 28.912 kg
Permissible ambient temperature 20 - 25 °C
Permissible relative humidity 80 %
Protection class according to DIN EN 60529 IP 20
RS 232 interface yes
Voltage 100 - 240 V
Frequency 50/60 Hz
Power input 120 W