Description

C 1 Package 2/12

The C 1 static jacket oxygen bomb calorimeter provides a high degree of automation and has the smallest calorimeter footprint available. It operates according to DIN 51900 and ISO 1928. The temperature readings are analyzed according to the classical isoperibol method of Regnault Pfaundler. The light and simply attachable combustion chamber allows an easy measurement preparation. Due to the variety of different interfaces (PC, balance, printer) this unit is easily adaptable depending on the customer’s specific application needs. Further adaption to data management and LIMS is possible with our calorimeter software C 6040 Calwin (accessory).

The C 1 calorimeter can only be operated together with a fitted cooling water supply unit. Therefore we recommend our C 1 Package 1/12 for best performance. It includes the chiller RC 2 basic and ensures best possible connectivity and function of the calorimeter.

The scope of delivery includes all parts necessary to set up the unit. Additionally it includes wear parts and consumables for the first approximately 500 experiments including 25 calibrations. The catalytic activated, halogen resistant combustion chamber C 1.12 is specifically designed for the combustion of samples that contain halogens and sulfur ( > 3 % ). It is equipped with the C 5010.5 large crucible holder and C 6 large quartz crucible.

Scope of delivery
  • C 1 S000
  • C 1.12 Combustion chamber, halogen resistant

Technical Data

Measuring range max. 40000 J
Measuring mode static jacket 22°C yes
Measuring mode static jacket 30°C yes
Measurements/h static jacket 4
Reproducibility static jacke (1g benzoic acid NBS39i) 0.15 %RSD
Working temperature 20 - 30 °C
Temperature measurement resolution 0.0001 K
Cooling medium temperature 18 - 29 °C
Cooling medium permissible operating pressure 1.5 bar
Cooling medium tap water
Type of cooling flow
Flow rate 50 - 60 l/h
Rec. flow rate at 18°C 55 l/h
Oxygen operating pressure max. 40 bar
Interface scale RS232
Interface printer RS232
Interface PC RS232
Oxygen filling yes
Degasification yes
Decomposition vessel integrated yes
Decomposition vessel, halogen resistant yes
Analysis according to DIN 51900 yes
Analysis according to ISO 1928 yes
Dimensions (W x H x D) 290 x 280 x 300 mm
Weight 22.59 kg
Permissible ambient temperature 5 - 40 °C
Permissible relative humidity 80 %
Protection class according to DIN EN 60529 IP 20
RS 232 interface yes
USB interface yes
Voltage 100 - 240 V
Frequency 50/60 Hz
Power input 120 W