What is USP 797, and why does one chapter of the United States Pharmacopeia decide how your cleanroom is built, how your staff dresses, and how long your preparations can be used? USP General Chapter 797 is the national standard for compounding sterile preparations, and state boards of pharmacy inspect against it in hospitals, 503A pharmacies, and clinics alike. It reaches from the ISO Class 5 laminar flow hood where you work to the sterile cleanroom gloves on your hands. This overview explains what the chapter covers, how it classifies preparations, and what the surrounding environment must provide.
Key Takeaways
- USP 797 sets minimum standards for preparing compounded sterile preparations (CSPs) in any setting, from hospital IV rooms to physician offices.
- USP writes the standard, but state boards of pharmacy and accreditors enforce it, sometimes with stricter local rules.
- The 2023 revision replaced low-, medium-, and high-risk levels with Categories 1, 2, and 3, and your category caps the beyond-use date you can assign.
- All sterile manipulations happen in an ISO Class 5 primary engineering control, and where that device sits determines your category.
- Personnel competency, cleaning frequency, and environmental monitoring carry as much weight in an inspection as the room itself.
- Category 3 buys longer BUDs in exchange for sterility testing, stability data, and tighter monitoring.
Products in This Article
The sterile compounding supplies featured in this guide are available at Lab Pro:
Texwipe Sterile VERTEX Prewet IPA Wiper, 12" x 12" (TX3042P) Pre-wetted sterile 70% IPA wipers that deliver the same alcohol load on every wipe of the ISO 5 work surface. |
Ansell DermaShield 73-711 Sterile Neoprene Gloves Sterile, powder-free neoprene gloves for aseptic manipulations inside the primary engineering control. |
Tians Bouffant Caps, 21" White Pleated Polypro (31588-21) Pleated polypropylene caps that cover all hair during the USP 797 garbing sequence. |
|
What Is USP 797?
USP 797 (formally USP General Chapter 797, Pharmaceutical Compounding: Sterile Preparations) is the United States Pharmacopeia standard that sets minimum requirements for compounding sterile preparations for human and animal patients. It covers personnel training, garbing, facility design, environmental monitoring, cleaning, beyond-use dating, and quality assurance, with one goal: preventing microbial contamination, excessive endotoxins, and errors in strength or ingredients.
As USP explains on its 797 overview page, medications that must be sterile include injections, IV infusions, and drugs given into the eye or the spine. The chapter applies to anyone who compounds them, whether that is a pharmacist, a nurse, or a physician.
What USP 797 Does Not Cover
Because the chapter is built around sterility, several activities fall outside it or are shared with other standards:
- Nonsterile compounding, such as capsules, creams, and oral liquids, is governed by USP 795.
- Hazardous drug handling adds USP 800 requirements on top of USP 797.
- Preparing a single dose of a conventionally manufactured product exactly as its approved labeling directs is not considered compounding.
- 503B outsourcing facilities follow federal cGMP instead of using USP 797 as their governing standard.
Why USP 797 Exists and Who Enforces It
USP writes standards; it does not enforce them. Knowing who does, and why the rules got stricter over time, tells you what an inspector will expect to see.
A Standard Shaped by Hard Lessons
USP 797 first became official in 2004 and was substantially revised in 2008. The current revision was published November 1, 2022, and became official November 1, 2023. The history behind it is sobering: a CDC update on the 2012 multistate outbreak tied to contaminated compounded methylprednisolone counted 753 cases of fungal infection. Outbreaks like that one are why the chapter leaves so little to interpretation today.
Who Enforces USP 797
Enforcement comes from the bodies that adopt the chapter:
- State boards of pharmacy, which inspect licensed pharmacies and may layer stricter requirements on top of USP
- Accreditation organizations, such as The Joint Commission for hospitals and ACHC for compounding pharmacies
- FDA, which can inspect compounders for cause and act against drugs prepared under insanitary conditions
Always check your own state's version. Some states adopted the 2023 revision as written, while others modified specific sections or set their own effective dates.
How USP 797 Categorizes Compounded Sterile Preparations
The 2023 revision dropped the old risk levels in favor of categories defined by the environment and the controls in place. Your category sets the maximum beyond-use date (BUD), which is the date or time after which a CSP cannot be stored or used.
Immediate-Use CSPs
Immediate-use CSPs are prepared outside a classified environment for urgent situations, such as an emergency department or a code. Administration must begin within 4 hours of preparation, and the preparation is limited to no more than three different sterile products. They are the exception, not a workaround for missing a cleanroom.
Categories 1, 2, and 3
| Category | Where It Is Prepared | Maximum BUD (Summary) | What It Requires |
|---|---|---|---|
| Category 1 | ISO 5 PEC in a segregated compounding area or a cleanroom suite | 12 hours at room temperature or 24 hours refrigerated | Core USP 797 requirements |
| Category 2 | ISO 5 PEC inside a cleanroom suite | About 1 to 90 days, depending on sterilization method, starting components, sterility testing, and storage | ISO 7 buffer room and ISO 8 ante-room |
| Category 3 | ISO 5 PEC inside a cleanroom suite with added quality controls | Up to 180 days, supported by stability data | Sterility testing, sterile garb, more frequent monitoring and personnel testing |
The ASHP summary of key USP 797 changes is a useful side-by-side reference if you are converting SOPs written under the 2008 risk levels.

Explore Lab Pro's full range of cleanroom consumables
Shop Cleanroom ConsumablesExplore Laminar Flow Hoods
Gloves, wipers, garb, and disinfectants chosen for ISO 5, 7, and 8 environments.
The Sterile Compounding Environment at a Glance
USP 797 builds contamination control in layers, with each space protecting the cleaner space inside it. This section is the overview; our USP 797 cleanroom requirements guide goes deeper on design and supply planning.
Primary Engineering Controls
The primary engineering control (PEC) is the ISO Class 5 device or zone where sterile manipulations happen. Common types include laminar airflow workbenches, biological safety cabinets, compounding aseptic isolators, and integrated vertical laminar flow zones. ISO Class 5 under ISO 14644-1 allows no more than 3,520 particles of 0.5 µm or larger per cubic meter of air, and critical manipulations must happen in first air, the HEPA-filtered air coming directly off the filter face.
Secondary Engineering Controls
The rooms around the PEC make up the secondary engineering control. For non-hazardous sterile compounding, the targets and the supplies each space runs on look like this:
| Space | ISO Class | Air Changes and Pressure | Recommended Products |
|---|---|---|---|
| Primary engineering control | ISO 5 | Unidirectional HEPA airflow; certified every six months | Laminar Flow Hoods, Sterile Gloves, Sterile IPA |
| Buffer room | ISO 7 | 30 total HEPA-filtered ACPH (at least 15 from the HVAC); at least 0.020 in. w.c. positive to the ante-room | Cleanroom Gowns, Sterile Wipers, Cleanroom Mops |
| Ante-room | ISO 8, or ISO 7 if it opens to a negative-pressure room | 20 ACPH; at least 0.020 in. w.c. positive to unclassified space | Sticky Mats, Masks, Shoe Covers |
| Segregated compounding area | Unclassified | No ISO class or pressure differential required; Category 1 only | Laminar Flow Hoods, Sterile Gloves |
Monitor pressure differentials continuously, and certify every PEC and classified room before use and at least every six months after that. USP also recommends keeping classified rooms at 20°C (68°F) or cooler with relative humidity below 60% to limit microbial growth and keep garbed staff comfortable.
People, Garbing, and Aseptic Technique
People are the largest source of contamination in any cleanroom, which is why USP 797 spends so much space on how compounding staff prepare, dress, and prove their skills.
Garbing Order and Components
Staff remove outer garments, jewelry, and cosmetics, then garb from dirtiest to cleanest. Category 3 compounding requires sterile garb with no exposed skin in the buffer room. Here is the standard sequence, with the matching supplies for each step:
1. Shoe Covers
Donned first, at the line between the dirty and clean sides of the ante-room.
2. Hair and Facial Hair Covers
Bouffant caps that cover all hair, plus beard covers where needed. Cleanroom-packed caps with welded seams shed the least.
Tians Bouffant Caps, 21" White Pleated Polypro
3. Face Mask
Covers the nose and mouth before hand hygiene, so you wash your hands last before gowning.
4. Hand Hygiene and Low-Lint Gown
Wash hands and forearms, then don a low-lint gown with sleeves that fit snugly at the wrist.
5. Sterile, Powder-Free Gloves
Applied after an alcohol-based hand rub, then sanitized with sterile 70% IPA throughout compounding.
Ansell DermaShield 73-711 Sterile Neoprene Gloves
Our cleanroom garbing guidelines walk through the full process, including the extra layers used for hazardous drug compounding.
Competency Testing
Garbing and aseptic technique must be demonstrated, not assumed:
- Initial qualification: hand hygiene and garbing competency with gloved fingertip and thumb sampling on three separate occasions, plus a media-fill test
- Ongoing requalification: at least every six months for Category 1 and 2 compounders, every three months for Category 3
- Cleaning competency: staff who clean and disinfect must be trained and evaluated too, not only those who compound
Cleaning, Monitoring, and Documentation
A certified cleanroom only stays compliant through routine cleaning, monitoring, and records. These are the areas where inspectors most often find gaps.
Cleaning and Disinfection Frequencies
| Area or Surface | Minimum Frequency | Recommended Products |
|---|---|---|
| PEC interior and equipment inside it | Daily on compounding days, plus sterile 70% IPA on the work surface during compounding | Sterile IPA, Sterile Wipers |
| Work surfaces outside the PEC, pass-throughs, floors | Daily on compounding days | Cleanroom Disinfectants, Cleanroom Mops |
| Walls, doors, ceilings, storage shelving | Monthly | Cleanroom Mops, Cleanroom Disinfectants |
| Sporicidal disinfectant | Monthly for Category 1 and 2; weekly for Category 3 | Cleanroom Disinfectants |
Inside the ISO 5 zone, your gloves, IPA, and wipers must be sterile. Pre-wetted sterile IPA wipers deliver the same amount of alcohol with every wipe, removing one variable from your daily PEC cleaning.
Texwipe Sterile VERTEX Pre-wetted IPA Wiper, 12" x 12" (TX3042P)
Get a cleaner ISO 5 work zone with fewer particles on every wipe, and fewer findings when you're inspected.
Environmental Monitoring
Monitoring proves the environment is doing its job between certifications:
- Viable air sampling: every six months for Category 1 and 2, monthly for Category 3
- Surface sampling: monthly for Category 1 and 2, weekly for Category 3
- Action levels in ISO 5: more than 1 CFU per cubic meter of air, or more than 3 CFU per surface sample, triggers investigation and identification to at least the genus level
- Daily checks: temperature, humidity, and pressure readings documented on compounding days
So, what is USP 797 in practice? It is the rulebook that ties every sterile preparation to a controlled environment, a qualified person, and a documented process. Your CSP category determines the space you need and the BUDs you can assign, while cleaning, monitoring, and competency testing keep that space compliant between certifications. Understand the structure first, then check your state's adoption of the chapter before you build SOPs around it.
Also, read:
- USP 797 Guidelines for Sterile Compounding
- Iso 5 Vs. Iso 7 Vs. Iso 8: Which Cleanroom Class Is Your Lab Operating In?
- 503A vs 503B Compounding Pharmacies: A Guide to Compounding Pharmacies
Partner With Lab Pro for USP 797 Compliance Supplies
At Lab Pro, our team is trained in USP 795, 797, and 800 requirements and supports sterile compounding pharmacies, hospital IV rooms, and outsourcing facilities with the supplies inspectors expect. We carry ISO 5 laminar flow hoods, sterile cleanroom gloves, low-lint PPE and safety apparel, and cleanroom consumables chosen for classified environments. When your garb, wipes, and disinfectants come from one source, your SOPs stay consistent from one order to the next.
Our Vendor Managed Inventory (VMI) program keeps those essentials stocked without adding work for your compounding team. We track usage, set par levels with you, and replenish on schedule, so a shortage of sterile IPA or gloves never delays a preparation. That leaves your staff free to focus on aseptic technique and patient safety.
Enhance your lab's efficiency and accuracy.
FAQs
Is USP 797 legally mandatory?
USP has no enforcement authority, so the chapter becomes mandatory when a regulator adopts it. Nearly every state board of pharmacy references USP 797 in its rules, and accreditors survey against it. In practice, any pharmacy or clinic compounding sterile preparations should treat it as required and confirm the exact version their state enforces.
Does USP 797 apply to physician offices and outpatient clinics?
Yes. The chapter applies to anyone and any setting that compounds sterile preparations, including physician practices, infusion centers, and surgery centers. Clinics that only prepare immediate-use CSPs or follow approved labeling for single doses have fewer requirements, but anything beyond those limits needs a compliant PEC and trained, qualified staff.
How is USP 797 different from USP 800?
USP 797 protects the preparation and the patient by keeping CSPs sterile. USP 800 protects workers and the environment from hazardous drugs. If you compound sterile hazardous drugs, you must meet both, which usually means a negative-pressure, externally vented buffer room with a containment PEC instead of the positive-pressure design used for non-hazardous work.
Can I use a compounding aseptic isolator instead of building a cleanroom?
An isolator in an unclassified, segregated compounding area can be used only for Category 1 CSPs with short BUDs. To assign Category 2 or Category 3 BUDs, the PEC must sit inside a cleanroom suite that meets USP 797 requirements. Isolators still need certification and routine cleaning like any other PEC.
What is a designated person under USP 797?
A designated person is the individual, or small group, assigned responsibility for the sterile compounding program. That includes overseeing training, monitoring results, SOP reviews, and corrective actions. Many pharmacies name the pharmacist-in-charge or a compounding supervisor, and some states require documenting the role in the pharmacy's policies.
How often does USP 797 get revised?
Major revisions are infrequent. The chapter changed significantly in 2008 and again with the version that became official in 2023. USP publishes proposed changes for public comment before they take effect, and state adoption can lag behind. Following USP's compounding updates and your board's newsletters helps you avoid surprises.